26
Connecting the High Voltage Test Lead
2.4 Connecting the High Voltage Test Lead
Connect the L2260 High Voltage Test Lead (option) to the voltage output terminal on the back of
the instrument.
The voltage output terminal on this instrument has a mechanism to prevent the LOW side lead from
coming off.
DANGER
Before using the instrument, check that the coating of the test leads are
neither ripped nor torn and that no metal parts are exposed.
Using the damaged test leads or instrument could result in serious bodily injury. If any
damage is found, replace the parts with those specified by Hioki.
Inspect the instrument and verify proper operation before use.
Use of the instrument while it is malfunctioning could result in serious bodily injury.
If you find any damage, contact your dealer or Hioki representative.
Never touch the test lead tip during testing (
[TEST]
lit).
Doing so could result in serious bodily injury or a short circuit.
Do not let the end of the test lead short-circuit two lines with voltage
applied.
A short circuit could result in serious bodily injury.
WARNING
Do not use test leads or the like whose insulation is damaged or whose
metal portion is exposed.
A hazardous voltage is applied to the measurement terminals. Doing so could result
in serious bodily injury.
Before connecting the L2260 High Voltage Test Lead and L2261
Unterminated Lead Cable, turn off the instrument and check that a high
voltage is not applied to the voltage output terminal.
•
The
DANGER
lamp is off.
•
[TEST]
is not lit in the display area.
Failure to do so could cause the operator to experience an electric shock or damage
the instrument.
Before starting a test, make sure that the L2260 High Voltage Test Lead
is properly connected.
Applying a voltage while the L2260 High Voltage Test Lead is not properly connected
could result in serious bodily injury.
When using the instrument, use the test leads specified by Hioki.
Using test leads other than those specified could cause bodily injury or short-circuit
accidents.
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...