213
Functional Specifications
Panel memory function
Operations
Data in memory
Function to save test condition settings in the instrument.
Test conditions can be loaded in the following ways.
• Instrument key operation
• Communications command through each interface
• Inputting of a signal to the external I/O terminal
DCW test
Test voltage, start voltage, test time, voltage rise time,
voltage drop time, judgment wait time, current judgment
upper-limit value, current judgment lower-limit value, arc
detection function, current compensation value of offset
canceling function, threshold value and compensation
value of contact check
IR test
Test voltage, test time, voltage rise time, voltage drop
time, judgment wait time, resistance judgment upper-limit
value, resistance judgment lower-limit value, resistance
compensation value of offset canceling function, threshold
value and compensation value of contact check
Program test
Test condtions and compensation value of each step,
number of test steps
BDV measurement
Test mode, initial voltage, end voltage, voltage rising
speed, voltage step, voltage holding time, number of steps,
arc detection function, distance between electrodes, upper
reference value, number of tests
Number of measured
values that can be
saved in the memory
DCW:
IR:
Program test:
BDV measurement:
Up to 64 patterns
Up to 64 patterns
Up to 50 steps, 30 patterns
Up to 10 patterns
Data memory function
Operations
Information to be
saved
Operation mode
Default setting
Test results (up to 32,000 results) can be saved in the
instrument.
Data can be loaded using a communications command or
USB flash drive.
No backup function (Data is deleted when the instrument is
turned off.)
Test mode: DCW, IR, W-IR/IR-W, program
Judgment result: pass, upper fail, lower fail, stop
Elapsed test time (sec.)
Voltage measured value
Current measured value
Measured resistance value
Test start time
On, off
Off
GFI function
Operations
Operation mode
Default setting
Function to protect operators from electric shock.
This function shuts down the output when a current over 0.5
mA flows from the high-voltage output part to the ground
during a test.
On, off
On
Buzzer
Operations
The buzzer volume for pass or fail judgment can be set.
The buzzer sounds approximately 50 ms when a pass
judgment is made and approximately 1 s when a fail
judgment is made.
Buzzer volume setting range for pass judgment: 0 to 5 (default setting: 3)
Buzzer volume setting range for fail judgment: 0 to 5 (default setting: 3)
Specifications
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...