Setting the Upper-Limit Value for Judgment
4.5 Setting the Upper-Limit Value for Judgment
The upper limit for the current value or resistance value is set as the reference for pass/fail
judgment.
When the measured current value or resistance value exceeds the reference value, a fail judgment
is determined for the test target.
If a pass/fail judgment is not to be performed based on the upper-limit value for an IR test, the test
upper-limit value can be set to off.
DCW setting
1
1
1
Press
[BASIC]
>
[UPPER]
on the setting screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.010 mA
to
100 mA
Default setting: 0.011 mA
The lower limit is set by the
[LOWER]
value for the
lower-limit value for judgment.
[LOWER]
<
[UPPER]
(Only when
[STATE]
is set to on)
IR setting
1
1
1
Press
[BASIC]
>
[UPPER]
on the setting screen.
2
2
3
3
2
Switch
[ON]
and
[OFF]
under
[STATE]
.
Default setting: OFF
3
When
[ON]
is set, press
[LIMIT]
.
Withstand V
oltage
Test/Insulation Resistance
Test Mode
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...