51
Setting the Voltage Drop Time
4.11 Setting the Voltage Drop Time
The time period from the end of a test with a pass judgment until the test voltage drops to around 0
V can be set.
When the contact check function is
[ON]
, the voltage drop time cannot be set. (p. 105)
DCW setting
1
1
1
Press
[BASIC]
>
[FALL TIME]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.1 s
to
300 s
,
OFF
*
1
*1. Default setting
IR setting
1
1
1
Press
[BASIC]
>
[FALL TIME]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.1 s
to
300 s
,
OFF
*
2
*2. Default setting
Withstand V
oltage
Test/Insulation Resistance
Test Mode
HIOKI ST5680A961-00