180
Internal Circuit Configuration
Electrical specifications
Connector
used
D-sub 37-pin female, #4-40 locking screw
Input
Photocoupler-isolated no-voltage-contact input (current sink and current source output compatible)
Input on
Residual voltage 1 V or less
Input off
Open (interrupting current 100 µA or less)
Output
Photocoupler-isolated open-drain output (non-polar)
Maximum load
voltage
30 V
Residual voltage 1 V or less
Maximum output
current
50 mA/channel
Power supply
output
Output voltage
Sink output: 5.0 V
±
10%
Source output: −5.0 V
±
10%
Maximum output
current
100 mA
Isolation
Isolated from protective ground potential and measurement circuit
Line-to-earth
voltage
50 V DC, 33 V AC rms, 46.7 V AC peak or less
Connection examples
Input circuit
Instrument
Instrument
Instrument
Instrument
Programmable
controller
Programmable
controller
Programmable controller
Connection with (negative common output)
Programmable controller
Connection with (positive common output)
Connection to the switch
Connection to the relay
Input
Input
Input
Input
Output
Output
Common
Common
ISO_COM
ISO_COM
ISO_COM
NPN
NPN
NPN
ISO_COM
PNP
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...