151
Self-test Function
I/O hander test
Checks whether the output signals from the external I/O terminal are normal, and whether the input
signals can be read normally.
1
1
1
Press
[SYS]
>
[TEST]
>
[EXEC]
of
[I/O HANDLER TEST]
on the measurement
screen.
2
2
2
Perform an output signal test (I/O OUT).
Implements output signal tests.
When a button is pressed, the signal is output from the
I/O output pin with the name of that button (on).
When a button that is outputting the signal is pressed,
no signal is output from the I/O output pin with the name
of that button.
3
3
3
Perform an input signal test (I/O IN).
The signal wire names for signals that are being input
(on) from among the input signals will be lit green.
Signal wires from which signals are not being input will
be grayed out (unavailable).
System Settings
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...