60
Starting and Ending a Test
The test ends and
[VOLTAGE ERROR]
is displayed in the following cases.
• When the output voltage does not reach
±
5% of the set test voltage
• When the voltage deviates from the set test voltage (
±
10 dgt) during the test and does not fall
within
±
10 dgt of the set voltage within 5 seconds
When the test is to be forcefully terminated
Press the
STOP
button to stop the voltage output. In this case, a pass/fail judgment is not made.
Operations during a test
The following is an example of the operation after a test starts.
DCW test (pass judgment)
No judgment in
the upper/lower-
limit value setting
Upper-limit value
judgment start
Lower-limit value judgment start
Test time
Rise time
Delay time
0.1 s
Test voltage
Start voltage
Pass judgment
Fall time
DCW test (fail judgment)
Discharge time
No judgment in
the upper/lower-
limit value setting
Upper-limit value
judgment start
Lower-limit value judgment start
Test time
Rise time
Delay time
0.1 s
Test voltage
Start voltage
Fail judgment
IR test (pass judgment)
Discharge
time
Judgment start
Test time
Rise time
Delay time
0.1 s
Test voltage
Pass judgment
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...