62
Starting and Ending a Test
For IR
4
Measured
resistance value
Displays the measured resistance value.
6
Current value
Minimum
resistance value
Displays the current value and minimum value of the measured resistance
value during a test.
DANGER
lamp
This is the lamp indicating a voltage is being output. It lights up during a test (including the voltage
rise time and voltage drop time). This lamp lights up if a voltage over the safe voltage (approx. 0.06
kV DC) remains in the output terminal even after the test ends.
External I/O
• The TEST signal is turned on at the same time when the
[TEST]
display lights up. In addition, the
H.V.ON signal is also turned on at the same time when the
DANGER
lamp lights up.
•
The TEST signal and H.V.ON signal are turned off at the same time when the
[TEST]
display and
the
DANGER
lamp is turned off.