Error Display
Errors during measurement
Error Display
Description
Ext. I/O output
Solution
DISCHARGE TIMEOUT
ERROR
Forced termination due to
incomplete discharge
ERR
Check the safety and then check the test
target, connection, and wiring.
CONTACT ERROR
Contact check error
CONT_ERR
Check the contact of the test target to the
High Voltage Test Lead. (p. 105)
VOLTAGE ERROR
•A voltage over the range
specified in the output voltage
monitoring function has been
detected.
Withstand voltage test:
±
(5% of s50 V)
Insulation resistance test:
±
(5% of s10 V)
•The voltage deviates from the
test voltage during the test
and does not fall within
±
10
dgt of the set voltage within 5
seconds.
ERR
Check the test target, connection, and
wiring.
V CIRCUIT ERROR
An error in the voltage detection
circuit power supply has been
detected.
ERR
Turn off the instrument and make a repair
request.
Error at the start of measurement
Error Display
Description
Ext. I/O output
Solution
DOUBLE ACTION
The instrument cannot start
test because the double action
function is activated.
−
Press the
STOP
button and then press
the
START
button within approximately
TIME SETTING ERROR
There is an error in the test time
setting.
−
In the W-IR/IR-W test mode or program
mode, the test time is set to
[CONT]
.
Correct the setting.
JUDGE SETTING ERROR
There is an error in the test
upper-limit or lower-limit setting
(p. 43).
−
The test upper-limit and lower-limit
values are reversed. Correct the setting.
The lower refernce value falls
out of the accuracy guarantee
range.
−
Set the lower reference value within the
accuracy guarantee range.
DELAY SETTING ERROR
There is an error in the
judgment wait time (delay time)
setting and/or the test time
setting.
−
Set values so that the sum of the voltage
rise time and test time is more than the
delay time. (p. 48)
VOLTAGE LIMIT ERROR
There is an error in the limit
voltage value setting (p. 40).
−
The limit voltage value is set less than
the test voltage value. Correct the value
so that the test voltage value is less
than or equal to the limit voltage value.
(p. 40)
POWER OVER ERROR
There is an error in the test
voltage setting and/or the test
upper-limit value setting.
−
The power value calculated from the
test voltage and test upper-limit value
is over the rated load capacity. Change
the setting so that the product of the test
voltage and the test upper-limit value
is equal to or more than the rated load
capacity (500 W).
Maintenance and Service
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...