215
Functional Specifications
Protection function
Operations
The instrument enters the protection state, the output is
shut down, and a test is stopped in the following cases. The
relevant message is displayed on the screen.
Protection display
Description
Unlock method
POWER SUPPLY ERROR
An error in the power supply
circuit has been detected.
Turn off the instrument.
FAN ERROR
A fan stop has been detected
Turn off the instrument.
LED ERROR
The
DANGER
lamp LED failure
has been detected.
Turn off the instrument.
INTERLOCK STATE
Input of an interlock signal has
been detected.
Short-circuit the interlock signal to ISO_
COM and press the
STOP
button.
REMOTE CONNECTOR
Connection/disconnection of the
external switch terminal has been
detected
After checking the test condition, press
the
STOP
button.
GROUND FAULT
The GFI function has been
activated.
After checking the test condition, press
the
STOP
button.
OVERLOAD
A current 1.5 to 3 times the
maximum rated current has been
detected.
After checking the test condition, press
the
STOP
button.
SUPPLY VOLTAGE FREQ
ERROR
The mains frequency is less than
45 Hz. Otherwise, it exceeds
65 Hz.
Turn off the instrument.
POWER TEMP ERROR
The temperature is abnormal.
Turn off the instrument.
SUPPLY VOLTAGE
ERROR
The output of the power
supply circuit dropped due to a
momentary power failure or other
cause.
Turn off the instrument.
HIGHLOAD POWER
OUTAGE
Voltage continues to be generated
in the state of a circuit error due
to a momentary power loss or
other cause.
Turn off the instrument.
OVERLOAD POWER
OUTAGE
A relatively longer voltage dip
occurs during voltage output
when the load is large.
Turn off the instrument.
OUTPUT TIME LIMIT
The test time has exceeded the
time rating.
• Press the
STOP
button.
• Leave the instrument for 30 minutes.
Specifications
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...