211
Functional Specifications
Gradual voltage rise test
Setting details
Test frequency
Initial voltage
Voltage holding time
for each step
Voltage steps
Maximum number of
steps
Arc detection
Distance between
electrodes
Upper reference limit
Number of tests
DC
0.25 kV, 0.50 kV, 1 kV, 2 kV, any voltage
Setting range: 0.010 kV to 7.999 kV
(Set resolution: 0.001 kV, default setting: 0.010 kV)
20 s (default setting), 60 s, 300 s, hold, any time
Setting range: 1 s to 300 s
(set resolution 1 s, default setting: 20 s)
When hold is selected, pressing the
START
button,
inputting an START signal to the external I/O terminal, or
sending a communication command when the specified
step is in the hold state proceeds to the next step.
10% of the initial voltage, 0.1 kV, 0.2 kV, any voltage
Setting range: 0.010 kV to 5.000 kV
(set resolution: 0.001 kV, default setting: 10% of initial
voltage)
20 steps (default setting: 10 steps)
Operation mode: Off Detection off
On After detecting arc, stops test.
(No arc cannot be detected during the
test-voltage rise time between steps.)
(default setting: Off)
Setting value: Test voltage fluctuation rate 1% to 50%
(Set resolution: 1%, default setting: 1%)
Setting range: 0.01 mm to 10.0 mm
(Set resolution: 0.01 mm, default setting: 0.01 mm)
Setting range: 0.010 mA to 100.0 mA
(Set resolution: 0.001 mA [0.010 mA to 9.999 mA]
0.01 mA [10 mA to 99.99 mA]
Default setting: 0.010 mA)
Setting range: 1 to 20 times (default setting: 1 time)
Measurement details
•Dielectric breakdown voltage (unit: kV)
Highest voltage at which no dielectric breakdown occurs in the test piece for the
specified holding time
The median, average, and standard deviation are displayed.
•Dielectric breakdown strength (unit: kV/mm)
Value obtained by dividing the dielectric breakdown voltage by the distance
between two electrodes used for a test
The median, average, and standard deviation are displayed.
Voltage measurement
accuracy
±
5% rdg
Dielectric breakdown
judgment procedure
Current upper-limit reference value setting, arc discharge detection
Graph display
The voltage value and measured current value of each test are displayed in
chronological order.
Specifications
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...