56
Offset Cancel (Open Compensation)
4.14
Offset Cancel (
Open Compensation)
The current that flows when the test lead or jig is open is measured in advance and is subtracted
from the test measurement result.
Valid test mode: DCW mode, IR mode, W-IR/IR-W mode, and program mode
1
1
2
2
1
On the setting screen, press
[BASIC]
>
[OFFSET C.]
to set the offset cancel function
to
[ON]
.
The offset cancel function is enabled and
[OFS]
is
displayed on the measurement screen during a test.
Default setting: OFF
2
Press
[< MEAS]
.
3
3
3
Press
[ADJ]
.
Preset test conditions
The set test conditions are displayed on the
compensation execution screen.
• The current compensation values are acquired based on the set test conditions.
•
In the W-IR/IR-W mode, a compensation cannot be executed. Execute a compensation in the
DCW mode and IR mode.