Offset Cancel (Open Compensation)
Acquiring the compensation value
The current compensation value is acquired using the same procedure for starting a normal test
with the open circuit without connecting the test target.
1
Press
[ADJ]
on the measurement screen and
start the test on the compensation execution
screen.
Acquired compensation values
When the test ends, the current compensation values
are acquired.
Acquired compensation values
The current compensation values can also be checked
on the measurement screen.
• Current compensation values are acquired at the following timings.
When the set test time ends
When the
STOP
button is pressed halfway
• When the test time is set to
[CONT]
, press the
STOP
button to acquire the current compensation
values once the measured value is stabilized.
•
Even if the instrument is turned off, the acquired current compensation values are backed up.
IMPORTANT
•
If the test voltage is changed in DCW or IR mode, the correction value is initialized and the
offset cancel function is turned off.
If you change the test voltage, acquire the correction value again.
• A current of up to 10 mA can be compensated. If a current exceeding the compensation range
is detected, a compensation failure error (ADJUST FAILED) occurs. Depending on the test
conditions, the test may be interrupted without waiting for the test time to elapse. When the test
is interrupted, the correction value cannot be acqired (except when the
STOP
button is pressed
during the test).
Withstand V
oltage
Test/Insulation Resistance
Test Mode
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...