49
Setting the Test Time
IR setting
1
1
1
Press
[BASIC]
>
[TEST TIME]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.1 s
to
999 s
,
CONT
Default setting: 1.0 s
The lower limit is set by
[RISE TIME]
(voltage rise time)
and
[DELAY]
(judgment wait time).
[DELAY]
<
[RISE TIME]
+
[TEST TIME]
(Only when
[DELAY]
is not set to off)
Withstand V
oltage
Test/Insulation Resistance
Test Mode