63
Checking Short Circuit before Applying the Set Voltage (Short circuit check function)
•
The short circuit check function is operated using 100 k
Ω
as a criterion.
Therefore, if equipment operating at 100 k
Ω
or less is measured, a short circuit
check error occurs.
•
When
[SHORT CHECK TIME]
is set to
[AUTO]
and the equipment to be
measured is short-circuited, it takes 0.5 s for a short circuit check and “SHORT
CHECK 0.500 s” is displayed.
Setting to MANUAL
1
Open the Settings screen.
Select
[SET]
.
2
Set
[SHORT CHECK]
to
[ON]
.
(When
[ON]
is selected, the
[SHORT CHECK TIME]
setting is
displayed below.)
Selection
[ON]
[OFF]
Return to the
Measurement
screen.
3
Set
[SHORT CHECK TIME]
to
[MANUAL]
.
Selection
[MANUAL]
Return to the
Measurement
screen.
10
9
8
7
6
5
4
3
2
1