59
Checking Faulty Contact and Contact Status (Contact check function)
2-terminal contact check function
If the insulation resistance of the equipment to be measured is within the
measurement range of the instrument, a 2-terminal contact check can be performed
by setting the upper and lower limit values of the comparator.
(Example) When the insulation resistance of the equipment to be measured is 100 M
Ω
Test voltage
500 V
Upper limit value
110 M
Ω
Lower limit value
90 M
Ω
Test duration
1 s
1
Set the test voltage to 500 V.
2
Set the lower limit value to 90 M
Ω
and the upper limit value to 110 M
Ω
.
3
Set the test duration to 1 s.
4
Connect the HIGH and LOW side test leads to the equipment to be
measured.
5
Start a test.
6
When the measured value is Over.F and the U.FAIL lamp turns on, the
contact may be faulty or the probe wires may be broken.
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.