
36
Setting the Range
3.2 Setting the Range
There are two range settings, auto and manual. For the manual range setting, the
range can be chosen from 2 M
Ω
, 20 M
Ω
, 200 M
Ω
, 2000 M
Ω
, and 4000 M
Ω
. Note,
however, that the ranges that can be set varies depending on the test voltage. (p. 153)
The auto range may require some time before a measured value is displayed as the
range has to shift after a test starts. To eliminate this time, set the range manually
according to the object to be measured.
Setting the manual range
1
Open the Settings screen.
Select
[SET]
.
2
Set
[RANGE]
to
[HOLD]
.
Selection
[HOLD]
Return to the
Measurement
screen.
3
Select the range.
Item selection
Select the range.
Return to the
Measurement
screen.
When analog output is used, 0 to 4 V is output from the analog output terminal according to
Содержание ST5520
Страница 2: ......
Страница 8: ...vi Contents ...
Страница 22: ...14 Usage Notes ...
Страница 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Страница 82: ...74 Initializing the System Reset ...
Страница 88: ...80 Deleting the Panel Data ...
Страница 158: ...150 Command Reference ...
Страница 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Страница 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Страница 197: ...13 09 ...
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......