108
Using the Switched Probe
7.8 Using the Switched Probe
The optional switched probe enables the start and stop operations of the ST5520/
ST5520-01 while the probe is held in hand.
Before a test, set the push switch mode to either trigger (
[TRIG]
) or continuous
(
[CONT]
). It is initially set to continuous mode (
[CONT]
).
1
Open the Settings screen.
Select
[SET]
.
2
Select the operation mode for
[EXT.SWITCH MODE]
.
Selection
[CONT]
[TRIG]
Return to the
Measurement
screen.
[CONT]
(Continuous mode)
A test is performed only when the push switch is pressed.
When the push switch is released, the test is terminated.
[TRG]
(Trigger mode)
Press the push switch to start a test. The test continues
even when the push switch is released. When the push
switch is pressed during the test, the test is terminated.
• For the operation of the ST5520 and ST5520-01 when the switched probe is
used, refer to “Enabled/disabled input list” (p. 163).
• The 9299 Switched Probe can be used as an ordinary probe when the switch
signal line plug is not connected to the EXT.SW terminal.
Содержание ST5520
Страница 2: ......
Страница 8: ...vi Contents ...
Страница 22: ...14 Usage Notes ...
Страница 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Страница 82: ...74 Initializing the System Reset ...
Страница 88: ...80 Deleting the Panel Data ...
Страница 158: ...150 Command Reference ...
Страница 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Страница 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Страница 197: ...13 09 ...
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......