152
Charge (measurement)
current
1.8 mA
Short circuit current
Less than 2.0 mA
Analog output
Not isolated against the power supply protective ground terminal
and isolated against the measurement terminals.
Output voltage: Switching between FULL and EACH
1.
FULL 0 to 4 V is output in all measured resistance ranges.
Measurement voltage Displayed resistance Output voltage
(DC)
25 V ≤ V < 100 V
0.000 M
Ω
to 200.0 M
Ω
0 V to 4 V
100 V ≤ V < 500 V
0.000 M
Ω
to 2000 M
Ω
0 V to 4 V
500 V ≤ V ≤ 1000 V
0.000 M
Ω
to 4000 M
Ω
0 V to 4 V
All measurement
voltages
Over. F
4 V
Under. F
0 V
2.
EACH 0 to 4 V is output according to each resistance range.
Resistance range
Displayed resistance Output voltage
(DC)
2 M
Ω
0.000 M
Ω
to 2.000 M
Ω
0 V to 4 V
20 M
Ω
1.90 M
Ω
to 20.00 M
Ω
0.38 V to 4 V
200 M
Ω
19.0 M
Ω
to 200.0 M
Ω
0.38 V to 4 V
2000 M
Ω
(100 V ≤ V < 500 V)
190 M
Ω
to 2000 M
Ω
0.38 V to 4 V
4000 M
Ω
(500 V ≤ V ≤ 1000 V)
190 M
Ω
to 4000 M
Ω
0.19 V to 4 V
All resistance ranges
Over. F
4 V
Under. F
0 V
Analog output
accuracy
±2% f.s.
Load resistance
1 M
Ω
or more
Содержание ST5520
Страница 2: ......
Страница 8: ...vi Contents ...
Страница 22: ...14 Usage Notes ...
Страница 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Страница 82: ...74 Initializing the System Reset ...
Страница 88: ...80 Deleting the Panel Data ...
Страница 158: ...150 Command Reference ...
Страница 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Страница 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Страница 197: ...13 09 ...
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......