Appx.
3
Output Voltage and Measurement Resistance
Appx. 3 Output Voltage and Measurement
Resistance
1200
1000
800
600
400
200
0
0
0.002 M
0.005 M
0.02 M
0.05 M
0.2 M
0.5 M
2 M
5 M
100 M
2000 M
4000 M
1000 V
500 V
200 V
100 V
Measured resistance [
Ω
]
Output voltage
[V]
Output voltage depends on measurement current 1.8 mA.
If the equipment to be measured carries 1.8 mA or more, the voltage does not rise to the set
output voltage.
10
9
8
7
6
5
4
3
2
1