81
360093-1/H
Reference information
9.3.1.2
Power consumption
The power consumption for a particular incubator is specified on the Product Specification label. This value
reflects the maximum number of agitators operating inside the incubator. Power consumption is measured
in full load Amperes.
Input voltage
Model
PC100h,
PC100i
PC900h,
PC900i
PC1200h,
PC1200i
PC2200h,
PC2200i
PC3200h,
PC3200i
PC4200h,
PC4200i
115 V
8.0 A
9.0 A
9.0 A
11.5 A
12.0 A
14.5 A
230 V
5.0 A
4.5 A
4.5 A
7.0 A
7.0 A
8.3 A
9.3.1.3
Circuit breakers
All 115 V countertop models have one circuit breaker. All 115 V floor models and all 230 V countertop
models have two circuit breakers. 230 V floor models do not have circuit breakers. The ratings are as follows:
Input
Voltage
Model
PC100h,
PC100i
PC900h,
PC900i
PC1200h,
PC1200i
PC2200h,
PC2200i
PC3200h,
PC3200i
PC4200h,
PC4200i
115 V
12 A
12 A
12 A
15 A + 2 A
15 A + 2 A
20 A + 5 A
230 V
6 A + 6 A
6 A + 6 A
6 A + 6 A
-
-
-
9.3.1.4
Fuses
All 230 V floor models have two fuses. All other models do not have fuses.
Fuses are 5 mm x 20 mm, Time Lag (Type T).
Input
Voltage
Model
PC2200h, PC2200i
PC3200h, PC3200i
PC4200h, PC4200i
230 V
10 A + 10 A
10 A + 10 A
10 A + 10 A
9.3.1.5
Load capacity for alarm contacts
The terminals on the remote alarm interface have the following maximum load capacity:
►
i.Series models: 0.5 A at 125 V ac; 1 A at 24 V dc
►
Horizon Series models: 0.25 A at 60 V ac; 0.25 A at 100 V dc
9.3.1.6
Maximum current draw for internal outlets
All models except the PC4200 models have one or more outlets inside the incubator. The maximum current
draw for each outlet is 0.5 A.