43
360093-1/H
Using the incubator
To disable the sound for all audible alarms
►
(PC4200h models) Using the alarm key, turn the key switch to Position 3.
►
(All other Horizon models) Using the alarm key, turn the key switch to the Off (
) position.
To disable the sound for only the motion alarm (PC4200h models only)
►
Using the alarm key, turn the key switch to Position 2.
6.5
Adjusting the screen contrast
NOTE:
This section only applies to i.Series models.
You can adjust the contrast between the text and the background.
To adjust the screen contrast
►
From the HOME screen, do one of the following:
►
To decrease the contrast one level, press the center button. The text appears lighter.
►
To increase the contrast one level, press the second button from the left. The text appears darker.
NOTE:
If a power failure alarm is active, the screen backlight is not illuminated to
conserve the charge for the backup batteries. During a power failure, pressing
the contrast buttons does not affect the contrast.
6.6
Rolling out shelves
The PC2200 and PC3200 models have roll-out shelves to provide easier access to agitators.
!
WARNING:
On PC3200 models, do not roll out more than one shelf at a time if they have
agitators on them. Having more than one shelf open at a time may cause the
incubator to roll forward and cause injury.
6.7
Storing items in the built-in agitator
NOTE:
This section only applies to PC4200 models, which have a built-in agitator. For
information about loading platelet bags in a Helmer agitator that is installed in
the incubator, refer to the Flatbed Platelet Agitator Operation Manual for your
agitator.
!
WARNING:
Follow all chemical handling and disposal requirements and procedures specified
by your organization. For more information, see Section 1, “Working safely.”
Before storing items in the incubator, be sure that the temperature is correct and stable. If you are using the
incubator for the first time, or after a power failure that causes the chamber to reach room temperature, let
the incubator stabilize at the setpoint temperature before storing items in it. Stabilization may take as long
as 30 minutes.