30
360093-1/H
Platelet Incubator Operation Manual
4.4.5
Changing the setting for the chart paper timer
NOTE:
A password is needed to complete this task. For more information about the
password, see Section 4.6, “Changing the password.”
If your incubator has a chart recorder, you can set a visual alarm to activate when the paper needs to be changed.
One piece of chart paper records temperatures continuously for seven days. The monitoring system has a
timer that activates an alarm exactly 6.5 days from when the timer is reset. You can reset, enable, or disable
the timer, but you cannot change the timer value.
The default setting set by the factory is Enabled. Available options are Reset, Enabled, and Disabled.
Resetting the timer also enables it. Selecting Enabled enables and resets the alarm if it was disabled.
Selecting Disabled disables the alarm.
To change the setting for the chart paper timer
1
Access the CONFIGURATION screen.
a
From the HOME screen, press the MAIN button. The MAIN screen appears.
b
Use the buttons to navigate to and select the Edit Configuration option. You are prompted to enter
your password.
c
Enter your password. If the password is correct, the CONFIGURATION screen appears with the
first option highlighted.
2
From the CONFIGURATION screen, use the buttons to navigate to and select the System Options
option. The SYSTEM OPTIONS screen appears with the Language setting highlighted.
3
Press the DOWN button until the Chart Paper Timer setting is highlighted.
4
Press the INC or DEC button until the new setting that you want to use appears.
5
Do one of the following to save the setting:
►
If you
selected Enabled or Disabled: Press the BACK button to return to the previous screen, or
press the HOME button to return to the HOME screen. The setting is saved.
►
If you
selected Reset:
a
Press the DOWN button. The CHART RECORDER PAPER screen appears.
b
Press one of the PAPER-CHANGED buttons (either the third or fourth button from the left).
The SYSTEM OPTIONS screen appears with Chart Paper Timer set to Enabled.
c
Press the BACK button to return to the previous screen, or press the HOME button to return
to the HOME screen. The setting is saved.
4.5
Enabling and disabling the temperature graph
NOTE:
A password is needed to complete this task. For more information about the
password, see Section 4.6, “Changing the password.”
The monitoring system has a real-time temperature graph which displays the readings of the upper
temperature probe for the past 24 hours of operation. This graph appears on the bottom of the HOME
screen whenever the buttons have not been pressed for one minute, and there are no active alarms. The
graph clears if a button is pressed or an alarm activates.
You can control whether the graph appears. The default setting set by the factory is Enabled. Available
options are Enabled and Disabled.
For more information about the temperature graph, see Section 6.2.5, “Understanding the temperature
graph and when it appears.”