Preparing the monitoring system on i.Series incubators
4.2
Viewing current settings
Through the monitoring system, you can view some of your current settings.
To view current settings
►
Access the VIEW CONFIGURATION screen.
a
From the HOME screen, press the MAIN button. The MAIN screen appears.
b
From the MAIN screen, use the buttons to navigate to and select the View Configuration option.
The VIEW CONFIGURATION screen appears with a list of current settings.
The settings that appear on the VIEW CONFIGURATION screen are as follows:
Setting
Description
Clock mode
The format used to display the time on the HOME screen and on the
Event Log
Date format
The format used to display the date on the HOME screen and on the
Event Log
Door Ajar Timeout
The number of minutes the door can be open before an alarm activates
Pwr Failure Timeout
The number of minutes the incubator can be without AC power
before an alarm activates
High Alarm Setpoint
The temperature at which the High Temperature Alarm should
activate
Low Alarm Setpoint
The temperature at which the Low Temperature Alarm should
activate
Cond. Alarm Setpoint
The temperature at which the Condenser Temperature Alarm should
activate
Alarm Volume
The volume for audible alarms
Alarm Pulse
The pattern for audible alarms
Chart Paper Timer
or
Chart Paper Days Left
The setting for the chart paper timer (if it is disabled), or the number
of days remaining before the chart paper needs to be changed (when
applicable)
Temperature Graph
The setting for the temperature graph feature (Enabled or Disabled)
Agitator Alarms
The settings for the motion alarm for each Helmer i.Series agitator
that is installed in the incubator, or the built-in agitator on PC4200i
models. All incubators are programmed to monitor three positions,
even though some models accommodate fewer agitators. Positions
that are not in use are set to OFF. Positions that are in use are set to
ON and display the number of minutes the agitator can be stopped
before an alarm activates.