51
360093-1/H
Maintaining the incubator
Task
Frequency
Quarterly
Annually
As needed
(Floor models only)
Clean the door gaskets.
►
For more information and instructions, see Section
7.3, “Cleaning the incubator.”
(PC4200 models only)
Check moving parts for the agitator for cleanliness and
wear. Lubricate moving parts. Replace parts if necessary
or when prompted by the agitation maintenance reminder.
►
For more information and instructions, see Section
7.4, “Maintaining moving parts on the built-in
agitator.”
Check the temperature calibration for the temperature
controller readout and change it if necessary.
►
For more information and instructions, see Section 7.6,
“Calibrating the temperature controller readout.”
(i.Series models only)
Check the temperature calibration for the i.Center
monitoring system readout and change it if necessary.
►
For more information and instructions, see Section
7.7, “Calibrating temperature readings on the
i.Center.”
Test the chamber temperature alarms.
►
For more information and instructions, see Section
7.8, “Testing chamber temperature alarms.”
Test the power failure alarm.
►
For more information and instructions, see Section
7.9, “Testing the power failure alarm.”
(i.Series models with installed agitators and PC4200h
models only)
Test the motion alarm.
►
For more information and instructions, see Section
7.10, “Testing the motion alarm.”
(i.Series models only)
Test the door open alarm.
►
For more information and instructions, see Section
7.11, “Testing the door open alarm.”
(i.Series models only)
Test the no battery alarm.
►
For more information and instructions, see Section
7.12, “Testing the No Battery alarm.”