22
360093-1/H
Platelet Incubator Operation Manual
4.1.2.1
Understanding options available from the MAIN screen
NOTE:
For a list of the options available on all the screens, see
Section 9.1, “i.Center
monitoring system screen reference.”
The options available from the MAIN screen are as follows:
Select this option
To do this
Event Log
Access the EVENT LOG screen
►
View historical information about alarms and
operational events
System Alarm Test & Status
Access the SYSTEM ALARM TEST & STATUS screen
►
Start or stop an automatic test for temperature alarms
►
View the number of days remaining before the paper for
the chart recorder needs to be changed
►
View the current status of the door (OPEN or CLOSED)
►
View the current condenser temperature
Edit Configuration
Access the CONFIGURATION screen (password required)
►
Change the language used for text
►
Change date and time information
►
Change temperature units
►
Change the volume and pattern for audible alarms
►
Enable or disable the chart paper timer
►
Enable or disable the temperature graph display
►
Change alarm-related setpoints and timers
►
Calibrate the temperature probe reading
►
Change certain settings to the default values set by the
factory
►
Change the password, preventing unauthorized changes
►
Configure the AgiTrak system to monitor agitation
View Configuration
Access the VIEW CONFIGURATION screen
►
View the date and time formats
►
View alarm-related setpoints and timers
►
View the volume and pattern for audible alarms
►
View the setting for the chart paper timer
►
View the setting for the temperature graph display
►
View the settings for agitator alarms
Product/Company Information
Access the INFORMATION screen
►
View the software versions for control and display
components of the monitoring system
►
View information to contact Helmer
i.Help
Access the i.Help screen
►
Access the on-board help system
►
View real-time and historical agitation information
►
Reset cycle counters for agitators