GE Analytical Instruments ©2009
98 of 236
DLM 90488-01 Rev. A
Scheduling Autoreagent Adjustments in On-Line Mode
If you operate in On-Line mode and use the Autoreagent feature, the
Schedule Reagent
value
determines how often the Analyzer repeats the preliminary measurement. Set this value by
following these steps:
1. Select the
Setup
tab. Make sure On-Line mode is selected.
2. Press the
Configure
button. Make sure the
Reagent
button is set to
Auto
.
3. Press the
Schedule Reagent
button. A number pad displays on the right side of the screen.
Enter a number to specify how often the Analyzer calculates the automatic reagent values;
the unit of measure for the value entered here is hours.
4. Press
Enter
to continue.
Setting Manual Flow Rates
Manually set reagent flow rates by following these steps:
1. Select the
Setup
tab.
2. Press the
Configure Mode
button.
3. Make sure the
Reagent
button is set to
Manual
.
4. Press the
Acid
button and use the keypad to enter a value. Press the
Enter
button to save
the value, or press
Cancel
to retain the current setting without making any changes.
5. Press the
Oxidizer
button and use the keypad to enter a value. Press the
Enter
button to
save the value, or press
Cancel
to retain the current setting without making any changes.
6. Press the
Menu
button to return to the main
Setup
screen.
Table 5: Recommended Oxidizer Flow Rates
TOC Concentration
Oxidizer Flow Rate
25 - 50 ppm
2.8 - 13.5 µL/min
10 - 25 ppm
1.4 - 7.5 µL/min
5 - 10 ppm
0.7 - 2.8 µL/min
1 - 5 ppm
0.7 - 1.4 µL/min
<1 ppm
0.0 - 0.7 µL/min
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...