GE Analytical Instruments ©2009
110 of 236
DLM 90488-01 Rev. A
5. Press the
Interval
button to print the average TOC value for the period of time specified.
The options are as follows:
Every Sample
,
Every Hour
,
Every 2 Hrs.
,
Every 4 Hrs.
, or
Every
8 Hrs.
Activating Binary Input
You can start and stop analysis with binary input to the Analyzer.
1. Select the
I/O
tab.
2. Press the
Binary Input
button.
3. Press the
Binary Input
button, and select
On
(or
Off
if you wish to disable binary input).
4. Press the
Start Level
button and select
High
or
Low
. This value corresponds to the signal
that will cause the Analyzer to start analysis. Confirm the correct value with your remote
operations center.
When DataGuard has been enabled,
Binary Input
is not available, to provide security to comply
with 21 CFR Part 11.
Setting Alarm Values
The Analyzer has two customizable alarm outputs that can be triggered if the measured TOC, TC, or
IC exceeds a set value or if a warning or error occurs. By default, both alarms are disabled. When
external alarms are connected to the Analyzer, you can set the alarm levels by following the steps
below. Instructions for wiring the alarms can be found in the “Installation” chapter on page 68.
1. Select the
I/O
tab.
2. Press the
Alarm 1
or
Alarm 2
button.
3. Press the
State
button to toggle the alarm output from
Off
to
On
.
4. Press the
Value
button to set the value that triggers the alarm. The options are:
•
TOC
,
TC
,
IC
— The alarm is triggered when a specified value is exceeded. If
you select a Value of
TOC
,
TC
, or
IC
, press the
Level
button to set the alarm
threshold.
•
Warn/Error
— The alarm is triggered when the Analyzer issues a warning or
error.
•
Error
— The alarm is triggered when the Analyzer issues an error.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...