GE Analytical Instruments ©2009
171 of 236
DLM 90488-01 Rev. A
Chapter 7. Maintenance
Replacing Consumables and Maintenance Items
To ensure optimum performance of the Analyzer, routinely replace Analyzer consumables and
maintenance items. Refer to Table 17 below for the recommended maintenance schedule.
Chemical reagents, the in-line filter, UV lamp, and pump tubing
must
be purchased from GE
Analytical Instruments. The use of reagents from other sources or failure to replace the reagents on
the prescribed replacement schedule will invalidate the Analyzer’s warranty.
If you need additional assistance, contact GE Analytical Instruments Technical Support at
800.255.6964. Training by a qualified service technician also can be provided.
Note that this chapter covers replacement instructions for all routine maintenance items that are
replaced on a regular schedule; instructions for non-routine maintenance items are provided in
those items’ packaging.
A one-year maintenance record log is provided for convenience to ensure that all required
maintenance is performed when necessary (see Table 18 on page 193).
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...