GE Analytical Instruments ©2009
145 of 236
DLM 90488-01 Rev. A
9. When prompted, remove the Calibration Blank from the IOS System, insert the TOC
Calibration Standard (as KHP), and press
Next
.
10. When prompted, remove the TOC Calibration Standard (as KHP) from the IOS System, insert
the IC Calibration Standard (as Na
2
CO
3
), and press
Next
.
11. After the IC Calibration Standard (as Na
2
CO
3
) has been analyzed, the calibration summary
screen is displayed. The Analyzer indicates if the calibration passed or failed based on the
following criterion:
• The % Diff for the IC standard is ±10% or less.
If this condition is satisfied, press
Apply
to accept the calibration and continue.
If this condition is not satisfied, press the
Cancel
button to reject the calibration. You may
need to perform the calibration procedure again. However, first consult the chapter called
“Troubleshooting” on page 195 to determine if there is a problem with the Analyzer.
Percent difference is calculated as follows:
The Expected TC Concentration equals the Certified TOC Concentration of the standard,
plus the measured TOC Concentration in the Calibration Blank, plus the Measured IC
Concentration of the standard.
Expected IC is the measured value, with the lamp off, from the TC channel for the IC
standard, using the new TC calibration. The summary screen also displays the measured
values using the existing calibration, the difference between measured and expected
values and the values that would be obtained using the new calibration values (adjusted
values).
Note:
If you want to manually calculate the calibration formulas for your records, you must print
the Data History and the Calibration Constants from the Analyzer before
and
after
performing the new calibration, in order to have access the values of the TC Slope and IC
Slope.
% Diff = Measured Concentration - Expected Standard Concentration
Expected Standard Concentration
-----------------------------------------------------------------------------------------------------------------------------------------------------------------
100%
×
TCexp = TO ICstd
ICexp = (TCr TOCoffset) New TCslope
Old TCslope
----------------------------------
×
TOC offset
–
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...