GE Analytical Instruments ©2009
106 of 236
DLM 90488-01 Rev. A
2. Press the
Analog Out
button.
• Press the
Min
button to set the minimum value (in ppb), corresponding to the minimum
analog voltage. Enter a number and press the
Enter
button to save the value, or press
Cancel
to retain the current setting without making any changes.
• Press the
Max
button to set the maximum value (in ppb), corresponding to the
maximum analog voltage. Enter a number and press the
Enter
button to save the
value, or press
Cancel
to retain the current setting without making any changes.
• Press the
Error
button to set a value when the Analyzer issues an error (1, 2.5, 4, or 20
mA, or No Change).
• Press the
Standby
button to set a value when the Analyzer switches out of analysis
mode into standby mode (1, 2.5, 4, or 20 mA, or No Change).
• Press the
Value
button to set the output value that will be sent to the analog output.
The values can be TOC, TC, or IC. Press the
TOC
,
TC
, or
IC
button and then press
Enter
to change the value, or press
Cancel
to retain the current setting without making any
changes.
• Press the
Protocol
button to set the analog output behavior when the Analyzer is
running system protocols. Press either the
Analysis Results
or
Standby
button.
• To calibrate the values output via the 4-20 mA outputs, see “Calibrating Analog Output
Values” below.
Calibrating Analog Output Values
Calibrate the 4-20 mA output values by following these steps:
1. Open the Analyzer door (900 On-Line) or access the terminal block on the back of the
Analyzer (900 Portable) and connect wiring from the 4-20 mA outputs on TB 3 to a digital
multimeter.
2. Select the
I/O
tab.
3. Press the
Analog Out
button.
4. Press the
Calibrate
button.
5. Press the
Adjust
button for the 4 mA voltage.
6. Press the
Increase
or
Decrease
button to change the value. Note that you can change the
rate at which the Analyzer applies the adjustment by changing the selection under the
Rate
button.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...