GE Analytical Instruments ©2009
81 of 236
DLM 90488-01 Rev. A
Once DataGuard is enabled, you will immediately be required to log in to the Analyzer with the
default Administrator User ID and Password:
User ID: ADMIN
Password: GEAI
Proceed to the section “Establishing a New Administrator Account for DataGuard” below to
complete the DataGuard activation.
After installation is complete, you can add unique user accounts to the Analyzer. Refer to Chapter 5,
"Password Protection and DataGuard" for more information.
Establishing a New Administrator Account for DataGuard
After you log in to the Analyzer for the first time after activating DataGuard, create a new
administrator-level account and inactivate the default administrator account to ensure Analyzer
security. (Refer to Chapter 5, "Password Protection and DataGuard" for complete DataGuard
information.)
1. Press the
Menu
button and select the
DataGuard
tab.
2. Press the
Add User
button.
3. Specify a new User ID for the administrator and press the
Enter
button.
4. Specify the Password for the administrator User ID and press
Enter
.
5. Verify the Password and press
Enter
.
6. Press the
User Level
button and select
Administrator
.
7. Press the
Password Expired
button and select
False
.
8. Press the
Back
button.
9. Press the
Logout
button. You will now use the new administrator account and delete the
default account.
10. Press the
Login
button and enter the new User ID and Password you just created.
11. Press the
Menu
button and select the
DataGuard
tab.
12. Press the
Edit User
button.
13. Use the arrow buttons to highlight the default administrator account and press
OK
.
14. Press the
User Status
button and select
Inactive
.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...