GE Analytical Instruments ©2009
152 of 236
DLM 90488-01 Rev. A
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. If you have an ICR unit, make sure the lever is in the
Bypass
position.
3. Remove the protective cap from the standards vials, and avoid touching the septa to
protect against introducing foreign particles.
4. Press the Menu button, select the
Maintenance
tab and press the
Suitability
button.
At this point, the prompts on the Analyzer’s display screen will help guide you through the
system suitability verification process.
5. Insert the Reagent Water (Rw) into the IOS System and press
Next
.
6. When prompted, remove the Reagent Water (Rw) from the IOS System, insert the 500 ppb
Sucrose Standard (Rs), and press
Next
.
7. When prompted, remove the 500 ppb Sucrose Standard (Rs) from the IOS System, insert
the 500 ppb Benzoquinone Standard (Rss), and press
Next
.
8. After the 500 ppb Benzoquinone Standard Solution (Rss) has been analyzed, the System
Suitability Values screen is displayed (see Figure 27). The Analyzer indicates if the system
suitability passed or failed based on the following criteria:
Acceptance criteria for USP System Suitability is response efficiency between 85% and
115%.
Response efficiency is calculated as follows:
Response Limit is calculated as follows:
Response Limit = Rs - Rw
9. Press either the
Apply
or
Cancel
buttons, depending on whether you want the new data to
be saved as the Analyzer’s response limit.
If you apply the Response Limit, this new value determines when to trigger the Analyzer’s
Response Limit alarm, when activated (see “Setting Alarm Values” on page 110).
10. Remove the vial from the IOS System and slide the door closed.
11. If you have an ICR unit, return the lever to the
Inline
position.
RE = (Rss - Rw)
(Rs-Rw)
----------------------------
100
×
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...