4. Press the
Contrast Adjust
button and use the keypad to change the contrast level of the
display. Press
Enter
to change the contrast to the new value, and then push
Menu
to exit.
Saving System Settings
You can archive settings to the USB port and restore them at a later time. This feature provides a
backup mechanism in the event that you need to return to previous settings, including calibration
and user-configurable settings. Save your Analyzer’s system settings as follows:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
USB I/O
button. Make sure that the USB flash memory drive is attached to the
USB port.
4. Press the
Save System
button.
To use the restore feature, be sure that the USB flash memory drive containing the previously
exported data is attached to the Analyzer’s USB port, repeat Steps 1 through 3 above, and select
Restore System
.
Programming the TOC Autozero
If your application requires you to perform on-line TOC measurements, you may want to set the
Analyzer to automatically perform a TOC Autozero at a specific interval. The TOC Autozero is
recommended only if the TOC concentration in the sample stream is 50 ppb or less. The TOC
Autozero function is described in more detail on page 140.
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Press the
Menu
button, select the
Maintenance
tab, and press the
Calibrate
button.
3. Press the
Program Autozero
button.
4. Press the
State
button and select
On
.
5. Press the
Frequency
button and use the number keypad to enter how often you want to
program the TOC Autozero function to work. The
Frequency
is measured in days; if you
enter 7 the TOC Autozero will run every week starting one week from the date you specify.
For example, if you enter 14, it will run every two weeks.
6. Use the date and time buttons to specify the start date and time for the TOC Autozero.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...