GE Analytical Instruments ©2009
197 of 236
DLM 90488-01 Rev. A
5 - UV Lamp Warning
Message: “The estimated lamp life is less than 15 days. Order a new UV lamp.”
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the UV lamp
life remains. Order a new UV lamp from GE now to avoid down time when the lamp expires.
6 - Tubing Warning
Message: “The estimated tubing life is less than 15 days. Order new tubing.”
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the sample
pump tubing life remains. Order new pump tubing from GE now to avoid down time when the
tubing expires.
7 - Resin Bed Warning
Message: “The resin bed life is less than 15 days. Order a new resin bed.”
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the resin bed
life remains. Order a new resin bed from GE now to avoid down time when the resin bed expires.
8 - User List Warning
Message: “The user list is nearly full. Archive the user list before adding more users.”
Explanation: DataGuard allows a maximum of 100 users. This warning displays when 90 user
accounts have been created.
9 - History Data Warning
Message: “The history data is nearly full. Archive the history data before taking more
measurements.”
Explanation: The data history can accommodate approximately 33,000 entries (about 90 days of
usage when the Analyzer is not in the optional Turbo mode). This warning displays when the data
history is 90% full (when approximately 31,941 entries have been saved).
10 - Audit Trail Warning
Message: “The audit trail is nearly full. Archive the audit trail before taking more measurements.”
Explanation: The audit trail can accommodate approximately 1,200 entries. This warning displays
when the audit trail is 90% full (when approximately 1,080 entries have been saved).
11 - Printer Timeout
Message: “Printer has timed out. Check that the printer is on and selected.”
Explanation: The printer did not respond and is not ready to print. Confirm that the printer’s
selection light is green (you may have to press the
SEL
or
Online
button on the printer).
12 - High IC
Message: “The ratio of TOC to IC is less than 0.1. You may need an ICR unit for this application.”
Explanation: This warning displays when the IC value is 10 times higher than the TOC value. If the
message persists, you may be experiencing high enough IC levels to merit using an ICR unit in your
configuration. This message also can be generated when analyzing samples with low
concentrations of TOC; if the TOC measurements are negative, you may need to perform a TOC
Autozero.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...