Figure 24: Reviewing TC and IC Values from a Single-Point Calibration
12. Remove the IC Calibration Standard from the IOS System.
13. You should now verify the calibration. Proceed to “Performing a Calibration Verification” on
page 149 for instructions.
Performing a Multi-Point Calibration
Before performing a multi-point calibration, save the current Analyzer system settings and evaluate
whether a TOC Autozero is required, as described under the section called “Preparing for
Calibration” on page 139.
When you are ready to calibrate the Analyzer, follow these steps:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. If you have an ICR unit, place the lever into the
Bypass
position. If you leave the lever in the
Inline
position, a warning message will display later.
3. Press the
Menu
button, select the
Maintenance
tab, and then press the
Calibrate
button.
4. Press the
Multi Pt Cal
button.
5. If you have not performed a reagent flush in the last 8 hours, you must perform one now.
Make sure the Analyzer is connected to a water supply and click
Yes
.
Cancel
Apply
Calibrate
12 Apr 2004 08:09
TC Calibration
Expected
Measured
Diff
Adjusted
Blank
N.A.
26.0ppb
N.A.
27.1ppb
25ppm
25.3ppm
24.9ppm
-1.58%
25.3ppm
IC Calibration
Expected
Measured
Diff
Adjusted
25ppm
26.4ppm
26.1ppm
-1.14%
26.4ppm
Calibration passed
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...