GE Analytical Instruments ©2009
175 of 236
DLM 90488-01 Rev. A
Replacing the Chemical Reagents
The oxidizer and acid reagents
must
be purchased from GE Analytical Instruments to ensure
product purity and operating life. The use of reagents from other sources or failure to replace the
reagents on the prescribed replacement schedule will invalidate the Analyzer’s warranty.
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. Turn off power to the Analyzer with the main power switch.
3. Open the Analyzer front panel by turning the two latches and swinging the panel door to
the left.
4. Locate the reagent enclosure, located at the top left-hand side of the Analyzer as you face
the Analyzer. Loosen the set screw and remove the reagent enclosure cover.
5. For each cartridge that you are replacing:
• Remove the cartridge from the enclosure.
• Push the red button to close the valve.
• Loosen the PEEK nut and disconnect the cartridge from the reagent syringe tubing.
Warning
Reagent containers are for single-use only:
do not refill
. Refilling or reusing
reagent containers will void all Analyzer and parts warranties and nullify any
performance claims.
Warning
To avoid exposure to the chemical reagents, wear acid-resistant gloves, protective
clothing, and safety goggles or a face shield when changing the reagent supplies.
Warning
Acid or oxidizer may spill from the containers if the valves are not closed.
Note:
Reagent containers need to be disposed of according to local regulations. All reagent
containers are shipped with Material Safety Data Sheets (MSDS). These sheets will
provide instructions. GE Analytical Instruments can not accept containers for disposal.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...