The same verification procedure is used whether a multi-point or single-point calibration was
performed. Verification occurs at a single concentration, which you can select. Choose a
concentration that is appropriate for your application. Verification can be repeated at multiple
concentrations if it is desirable to verify the entire analytical range of the instrument instead of
verification near the range of interest only.
Follow these steps to perform a calibration verification:
1. If the Analyzer is taking measurements, press the
Stop Analysis
button.
2. If you have an ICR unit, make sure the lever is in the
Bypass
position. If you leave the lever
in the
Inline
position, a warning message will display later.
3. Press the
Menu
button, select the
Maintenance
tab, and then press the
Calibrate
button.
4. Press the
Verify
button.
5. If you are performing the verification immediately after the calibration, press the
Skip
button to avoid flushing the reagent syringes. If there has been at least an eight-hour delay
between performing the calibration and starting the verification, press
Flush
to perform a
reagent flush now.
Make sure the Analyzer is connected to a water supply and click
Yes
.
If you have performed a reagent flush in the last 8 hours, click
No
to skip the flush and
proceed to Step 6.
6. When the flush is complete, click
OK
to continue. Remove the protective cap from the
standards vials, and avoid touching the septa to protect against introducing foreign
particles.
7. Press the
Ver. Standard
button, select the concentration of the standard you will be
measuring, and then push
Next
.
At this point, the prompts on the Analyzer’s display screen will help guide you through the
verification process.
8. Insert the Verification Blank into the IOS System and press
Next
.
9. When prompted, remove the Verification Blank from the IOS System, insert the TOC
Verification Standard (as sucrose), and press
Next
.
10. When prompted, remove the TOC Verification Standard (as sucrose) from the IOS System,
insert the IC Verification Standard (as Na
2
CO
3
), and press
Next
.
11. After the IC Verification Standard (as Na
2
CO
3
) has been analyzed, the Verification summary
screen is displayed. The Analyzer indicates if verification passed or failed based on the
following criteria:
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...