GE Analytical Instruments ©2009
200 of 236
DLM 90488-01 Rev. A
flash memory drives, follow the Sievers 900 TOC Analyzers link on our Web page,
www.GEInstruments.com.
50 - Unstable Measurements Error
Message: “Unstable measurements detected during calibration/verification. Please refer to the 900
Oper. and Maint. Manual for recommendations.”
Explanation: This error displays when the results from calibration/verification are inappropriate and
calibration/verification cannot continue. Follow the suggestions in the section called “Step 2: Visual
Inspection” on page 201 before continuing with analysis. You may need to contact Technical
Support to confirm that the Analyzer is functioning properly.
51 - History Flash Error
Message: “A problem has been detected in the history data flash. Please contact GE.”
Explanation: The Analyzer experienced a problem when trying to store data in the flash memory. It
is possible a section of the flash memory is damaged. To protect against losing data, stop analysis
and contact Technical Support.
52 - User List Error
Message: “The user list is full. Archive the user list before adding more users.”
Explanation: DataGuard allows a maximum of 100 users. This error displays when 100 user
accounts have been created. If you have inactive accounts in the user list, you can archive users
(
DataGuard
tab
Archive Users
) to remove them from the list; you will then be able to add
additional users.
53 - History Data Error
Message: “The history data is full. Archive the history data before taking more measurements.”
Explanation: This error displays when the data history is completely filled. Export (
Data
tab
Export
) or Print (
Data
tab
the data) before proceeding.
54 - Audit Trail Error
Message: “The audit trail is full. Archive the audit trail before taking more measurements.”
Explanation: This error displays when the audit trail is completely filled. To comply with 21 CFR Part
11, you must either Export (
DataGuard
tab
Export AuditTrail
) or Print (
DataGuard
tab
Audit Trail
) the audit trail before proceeding.
55 - ICR Error
Message: “A problem has been detected with the ICR. Check the ICR indicator lights. Refer to the
900 ICR Manual for recommendations.”
Explanation: This error generally displays when the ICR is not creating a vacuum but is set to Inline.
Refer to the 900 ICR
Operation and Maintenance Manual
for troubleshooting suggestions.
56 - Acid Syringe Error
Message: “Error detected with the acid syringe. Inspect the syringe and cables. Refer to the 900
Oper. and Maint. Manual for recommendations.”
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...