GE Analytical Instruments ©2009
85 of 236
DLM 90488-01 Rev. A
1. Select the
I/O
tab and press the
Printer
button.
2. Press the
Printer
button and select your printer model, either
Citizen
or
Seiko
. If you do
not have a printer, make sure
No Printer
is selected.
3. Press the
Header Freq.
button and select
First Page
to print a header only on the first
page of output or select
All Pages
to print a header on all pages of output.
4. Press the
Print Freq.
button and select how often you would like to print TOC data
information.
Exporting and Printing Constants
Calibration constants, reagent flow rates, and other key parameters are stored in the Analyzer’s
memory. You should export and print the factory settings for future reference.
To export the settings to a comma-separated text file, follow these steps:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
USB I/O
button. Make sure that the USB flash memory drive from the Analyzer’s
accessory kit is attached to the USB port, then press the
Save System
button. Archive the
exported files in a secure location on your computer.
4. Press the
Back
button and then press the
Menu
button.
If you have a printer attached to the Analyzer, you can print these settings for future reference by
following these steps:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
Print Constants
button.
4. Press each of the four buttons, one at a time, to print the different constants.
Set Up Data I/O
If you installed binary input, alarms, or the 4-20 mA output, configure settings for those features
now. Select the I/O tab and configure the appropriate settings. Note that you may need to consult
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...