Appendix
GE Analytical Instruments ©2009
229 of 236
DLM 90488-01 Rev. A
• The location of HyperTerminal varies from computer to computer. Often it can be
launched by exploring the Start menu path as follows:
Start
Programs
Accessories
Communications
HyperTerminal
.
• Select
Start
Run, type
Hypertrm
in the
Open
field, and click
OK
.
3. When HyperTerminal launches, the
Connection Description
window displays. Enter a
name in the Name field, such as
Sievers 900 TOC Analyzer
, and click
OK
.
4. When the
Connect To
window displays, select
Communications Port
from the
Connect
Using
drop-down menu. Click
OK
. If more than one communications port is in the list and
you’re not sure which one to select, see “Determining Your Communications Port Number”
on page 230.
5. In the
Com Settings
window, specify the following settings:
Bits Per Second:9600
Data Bits:8
ParityNone
Stop Bits:1
Flow Control:Hardware
After selecting these values, click
OK
.
6. The main HyperTerminal window displays. Select
Transfer
Capture Text
. Specify a
location and name for the file and click
OK
. Include a file extension of either
.txt
or
.csv
. For example,
JulyData.csv
.
7. On the Analyzer, export the data you want to capture. Be sure to select
Serial
as your
export destination.
8. When the export begins, you will see the data streaming in the HyperTerminal window.
When the export is complete (activity stops in the HyperTerminal window), select
Transfer
Capture Text
Stop
.
9. The data is now saved in the file location you specified in Step 6. When you open the file in
a spreadsheet or database program, you will need to browse to this folder in the program’s
Open
dialog box.
10. Exit out of HyperTerminal. You will be prompted to save the connection information. Saving
the information makes it easier to connect to the Analyzer the next time you want to
capture data.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...