GE Analytical Instruments ©2009
130 of 236
DLM 90488-01 Rev. A
• Press the
Min Length
button to set the minimum character length for a Password. The
default value is
3
characters.
• Press the
Login Timeout
button to set the number of minutes of allowable keypad
inactivity before users are required to re-enter their login information. The default value
is
5
minutes.
Changing User Passwords
All users can change their own Passwords by following these steps:
1. Log in with a valid User ID and Password.
2. Select the
DataGuard
tab.
3. Press the
Change Password
button.
4. Enter the old Password and press
Enter
.
5. Specify a new Password and press
Enter
.
6. Verify the new Password and press
Enter
.
Administrators can change the Password associated with any User ID by pressing the
Password
button on the
Edit User
screen (“Editing User Information” on page 128).
Dealing with Forgotten Passwords
If users forget their Passwords, the administrator must assign new Passwords by following these
steps:
1. Log in to the Analyzer with an administrator-level User ID and Password.
2. Select the
DataGuard
tab.
3. Press the
Edit User
button.
4. Select the appropriate User ID from the list using the scroll buttons and press
OK
.
5. Press the
Password
button.
6. Specify a new Password and press
Enter
.
Note:
Require a new user to change the Password upon log in so that it will be easy to
remember and unknown to the administrator.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...