GE Analytical Instruments ©2009
189 of 236
DLM 90488-01 Rev. A
changed is to periodically measure the flow rate of water out the waste line from the IOS System
and replace the filter element when the flow rate starts to decrease.
It is desirable to replace the filter element on a routine basis to prevent clogging. For this procedure,
two 3/4" wrenches are needed. To replace the filter element, follow these steps:
1. Stop the Analyzer by pressing the
Stop Analysis
button.
2. Turn off the Analyzer using the main power switch.
3. Shut off the water to the sample inlet system.
4. Remove the filter by loosening the Swagelok nuts on the 1/4" Teflon tubing and
disconnecting the tubing.
5. Position the 3/4" wrenches on the ends of the filter (see Figure 37). Loosen the adapter on
the inlet side of the filter.
6. Unscrew the spring-loaded inlet adapter, taking care not to lose the spring.
7. Remove the old filter element from the body of the filter.
8. Insert a new filter element into the body of the filter, opened end first.
9. Replace the spring in the inlet adapter and screw the inlet adapter into the body of the
filter.
10. Secure the inlet adapter by tightening approximately one-quarter turn past finger-tight.
11. Reconnect the 1/4" Teflon tubing from the sampling port adapter to the inlet of the in-line
filter. The in-line filter has arrows on the body of the filter that indicate the direction of flow
through the filter. Tighten the nut one-quarter turn past finger-tight.
12. Reconnect the 1/4" Teflon tubing to the outlet of the filter and tighten the nut one-quarter
turn past finger-tight.
13. Turn on the water supply at the sampling port.
14. Turn on the Analyzer using the main power switch.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...