GE Analytical Instruments ©2009
191 of 236
DLM 90488-01 Rev. A
• If you choose
New Oxidizer
, enter the expiration date that was listed on the oxidizer
packaging: press the
Day
button and use the number pad to specify the day of
expiration and press
Enter
; repeat for the
Month
and
Year
buttons. Press the
Amount
button and confirm that the correct volume is selected, either
150 mL
or
300 mL
. Press
Accept
after you have entered the complete expiration date. You will be prompted to
confirm the installation of new oxidizer; press the
Confirm
button to continue.
4. Repeat Step 3 for each consumable that you installed.
Figure 38: The Consumables Screen
Flushing the Reagent Syringes
If the Analyzer has not been used for over 24 hours, the decomposition of persulfate can produce
oxygen bubbles in the syringe pump and the reagent addition lines. To prevent bubbles from
entering the sample stream and interfering in the TOC measurement, a reagent flush is used to
remove the bubbles and fill the syringe with fresh reagent. A reagent flush also is recommended
after installing new chemical reagents.
For the reagent flush, it is recommend that the Analyzer must be connected to a water supply. A 40-
mL vial of water may be used for the reagent flush. To initiate a syringe flush, follow these steps:
Note:
When Password protection or DataGuard is activated, not all users are able to modify
settings on this screen. See “Using Password Protection” on page 125 or “Using
DataGuard” on page 126 for more information.
Consumables
4 Mar 2004 08:25
Main
Start Analysis
Lamp:
26 Days Used
176 Days Remaining
Tubing:
26 Days Used
296 Days Remaining
R. Bed:
26 Days Used
301 Days Remaining
Acid:
Expires 23 FEB 2005
357 Days Remaining
Oxid:
Expires 23 MAY 2004
85 Days Remaining
New Tubing
New R. Bed
New Acid
New Oxid
New Lamp
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...