10. After 10 minutes, open the Analyzer case and check the tubing that leads to the resin bed
(ion exchange column). Most of the air originally in the tubing should be replaced with
water. If more than a few small bubbles remain in the tubing, repeat Steps 6 through 9.
Figure 39: The DI Water Loop
To troubleshoot a lack of flow through the Analyzer, follow the steps in the next three sections:
“Checking the Sample Pump,” “Checking the Inlet Tubing,” and “Backflushing the Analyzer.”
Checking the Sample Pump
While the Analyzer is operating, open the cover and make sure the sample pump is turning. If the
pump is not turning and the Analyzer is operating (the display screen indicates that analysis is
taking place), contact GE Analytical Instruments Technical Support for instructions.
Warning
To avoid potentially dangerous shock, do not touch anything inside
the Analyzer while observing the sample pump.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...