Turbo Mode Specifications
When operated in Turbo mode, the Analyzer has the following specifications:
Linear range
0.20 ppb – 10 ppm TOC
Precision
2% RSD
Accuracy
±3%
Analysis Time:
4 seconds
Sample Flow Rate:
1.1 mL/min; 0.5 mL/min when TOC > 5 ppm
Instrument Delay (90% recovery)
3.5 minutes
Enabling Turbo Mode
Activation of Turbo sampling mode may be purchased directly from GE Analytical Instruments.
1. Select the
Setup
tab.
2. Press the
Enable Turbo
button.
3. Enter the Activation Code. If necessary, press the
Number
button to toggle the keypad
from letters to numbers.
4. Press the
Enter
button.
5. After Turbo mode has been enabled, you must turn Turbo mode on by following the
instructions in the next section.
Turning Turbo Mode On and Off
To turn Turbo mode on or off, follow these steps:
1. Press the
Menu
button and select the
Setup
tab.
2. Press the
Configure
button.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...