GE Analytical Instruments ©2009
196 of 236
DLM 90488-01 Rev. A
2. Press the
Warnings/Errors
button.
3. Use the scroll buttons to browse the warnings and errors.
4. Press the
button to print the list.
You also can print or export a full list of warnings and errors that includes the time each event
occurred, even after you have removed items from the Warnings/Errors list. To do so, follow these
steps:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
Warn/Error Out
button.
4. Press either the
Print History
or
Export History
button.
If you select
Export History
, you will be prompted to select the port for export. Press either
Serial
or
USB
to export the list of warnings and errors.
A complete list of warnings and errors is shown below.
Warnings and Error Messages
2 - Analog Warning
Message: “An analog error was detected. Turn power off and on to reset. Contact GE if the error
persists.”
Explanation: This warning generally displays when the temperature or conductivity for the
measurement is out of range. Turn the Analyzer off and on and restart analysis. If the warning
persists, contact Technical Support.
3 - Acid Warning
Message: “The estimated amount of acid is less than 10%. Confirm the amount is low. Order a new
acid container.”
Explanation: Based on the flow rate and history of usage, the Analyzer estimates that only 10% of
the acid reagent is remaining. Order a new acid container from GE now to avoid down time when
the acid is exhausted.
4 - Oxid Warning
Message: “The estimated amount of oxidizer is less than 10%. Confirm the amount is low. Order a
new oxidizer container.”
Explanation: Based on the flow rate and history of usage, the Analyzer estimates that only 10% of
the oxidizer reagent is remaining. Order a new acid container from GE now to avoid down time
when the oxidizer is exhausted.
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...