GE Analytical Instruments ©2009
97 of 236
DLM 90488-01 Rev. A
If you know the specific flow rates you want to use for each reagent, you can manually set the flow
rates, according to Table 5 on page 98 and Table 6 on page 99.
Using the Autoreagent Function
Set up the Autoreagent function by following these steps:
1. Select the
Setup
tab.
2. Press the
Configure Mode
button (see Figure 13).
3. Make sure the
Reagent
button is set to
Auto
. The Analyzer will automatically calculate the
flow rates for both reagents.
4. To modify the automatically calculated acid flow rate, press the
Acid
button and select an
adjustment value:
+10%
,
+20%
,
-10%
, or
-20%
. You also can return the value to
Auto
.
5. To modify the automatically calculated oxidizer flow rate, press the
Oxidizer
button and
select an adjustment value:
+10%
,
+20%
,
-10%
, or
-20%
. You also can return the value to
Auto
.
Figure 13: Selecting the Reagent Flow Method
Note:
Analysis must be stopped prior to changing the reagent flow rates.
Menu
Auto
Manual
Acid
0.05
Oxidizer
0.02
Rejects
1
Reps
5
Reagent
Manual
Configure
4 Mar 2004 08:16
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...