GE Analytical Instruments ©2009
207 of 236
DLM 90488-01 Rev. A
Figure 41: Comparing the Flow Rate Through the Sample Pump Tubing
5. Reconnect the pump tubing to the Y-fitting.
6. Follow the piece of tubing with the obstruction (the one with the slower flow) around the
pump and up to the green restrictor tubing. Disconnect the restrictor tubing from the
sample pump tubing.
7. Attach the water bottle to the green restrictor tubing, using a small length of tubing as a
connector (see Figure 42).
Figure 42: Flushing the Sample Side
Содержание Sievers 900 Series
Страница 10: ...GE Analytical Instruments 2009 10 of 236 DLM 90488 01 Rev A ...
Страница 64: ...GE Analytical Instruments 2009 64 of 236 DLM 90488 01 Rev A Figure 3 Left Side Dimensions of the Analyzer ...
Страница 88: ...GE Analytical Instruments 2009 88 of 236 DLM 90488 01 Rev A ...
Страница 122: ...GE Analytical Instruments 2009 122 of 236 DLM 90488 01 Rev A ...
Страница 159: ...GE Analytical Instruments 2009 159 of 236 DLM 90488 01 Rev A ...
Страница 161: ...GE Analytical Instruments 2009 161 of 236 DLM 90488 01 Rev A ...
Страница 165: ...GE Analytical Instruments 2009 165 of 236 DLM 90488 01 Rev A ...
Страница 167: ...GE Analytical Instruments 2009 167 of 236 DLM 90488 01 Rev A ...
Страница 170: ...GE Analytical Instruments 2009 170 of 236 DLM 90488 01 Rev A ...
Страница 172: ...GE Analytical Instruments 2009 172 of 236 DLM 90488 01 Rev A Figure 30 Interior Overview of the Analyzer ...
Страница 194: ...GE Analytical Instruments 2009 194 of 236 DLM 90488 01 Rev A ...
Страница 217: ...GE Analytical Instruments 2009 217 of 236 DLM 90488 01 Rev A Figure 47 Draining the Measurement Module ...
Страница 218: ...GE Analytical Instruments 2009 218 of 236 DLM 90488 01 Rev A ...
Страница 235: ...Notes GE Analytical Instruments 2009 235 of 236 DLM 90488 01 Rev A 236 ...
Страница 236: ...Notes GE Analytical Instruments 2009 236 of 236 DLM 90488 01 Rev A 236 ...