![GE AKTA ready Скачать руководство пользователя страница 79](http://html.mh-extra.com/html/ge/akta-ready/akta-ready_operating-instructions-manual_128010079.webp)
It is possible to assemble a flow kit before it is mounted onto the system cabinet. The
recommended procedure, however, is to connect the three parts to each other in con-
nection with mounting of the flow kit. This is the procedure described in this chapter.
Flow kit types
Two different ÄKTA ready flow kits are available:
•
ÄKTA ready High Flow Kit, with tubing diameter 9.5 mm (inlet manifold 12.7 mm,
pump tubing 12 mm). Intended for 10 to 20 liter ReadyToProcess columns.
•
ÄKTA ready Low Flow Kit, with tubing diameter 6.4 mm (inlet manifold 9.5 mm, pump
tubing 8 mm). Intended for columns with a volume of less than 10 liters.
Both flow kits use 25 mm TC connectors for connecting to inlet and outlet containers,
pump tubing, and column.
NOTICE
Only use flow kits, supplied by GE Healthcare.
Specifications for ÄKTA ready flow kits are found in
ÄKTA ready flow kit specifications,
. Ordering information is in
ÄKTA ready flow kits, on page 210
.
ÄKTA ready Operating Instructions 28960345 AD
79
6 Flow kit installation
6.1 ÄKTA ready flow kits